PULSe Trigger Commands for DS1000B Series Digital Oscilloscope

DS1000B Series Digital Oscilloscope covers following oscilloscopes:

  • DS1204B | 200 MHz Digital Oscilloscope
  • DS1104B | 100 MHz Digital Oscilloscope
  • DS1074B | 70 MHz Digital Oscilloscope

PULSe Trigger Commands

  1. :TRIGger:PULSe:MODE

Command Format:
:TRIGger:PULSe:MODE <mod>
:TRIGger:PULSe:MODE?
Function:
The command sets the trigger condition. <mod> can be: +GREaterthan (positive pulse width greater than), +LESSthan (positive pulse width less than), + EQUal (positive pulse width equal), -GREaterthan (negative pulse width greater than), -LESSthan (negative pulse width less than) or –EQUal (negative pulse width equal).
Returned Format:
The query returns +GREATER THAN or +LESS THAN, +EQUAL, -GREATER THAN, -LESS THAN, –EQUAL.
Example:
:TRIG:PULS:MODE +GRE                     Set the trigger condition as +GREaterthan.
:TRIG:PULS:MODE?                             Return +GREATER THAN.

  1. :TRIGger:PULSe:WIDTh

Command Format:
:TRIGger:PULSe:WIDTh <wid>
:TRIGger:PULSe:WIDTh?
Function:
The command sets the pulse width. The <wid> range is 20ns ~10s.
Returned Format:
The query returns the value of pulse width, and the unit is s.
Example:
:TRIGr:PULS:WIDT 0.001                                                 Set the pulse width as 1ms.
:TRIG:PULS:WIDT?                                            Return 1.000e-003.

All Trigger Control commands describe in
Trigger Control for DS1000B Series Digital Oscilloscope

All programming commands describe in
Command Systems for DS1000B Series Digital Oscilloscope