3.0 GHz Spectrum Analyzer GSP830 User Manual

3.0 GHz Spectrum Analyzer GSP830 User Manual
Pages 181
Views 1,524
Downloads : 1 times
PDF Size : 4.4 MiB

Summary of Contents

3.0 GHz Spectrum Analyzer GSP830 User Manual

  • MARKER 63 Activate/de-activate marker Range Normal marker 5 Delta marker 5 pairs Amplitude −120 to +20dBm, 0.1dB resolution −73.01 to +66.99dBmV, 0.01dB resolution −13.01 to +126.99dBuV, 0.01dB resolution Frequency 0kHz to 3GHz Activate normal marker 1. Press the Marker key. Marker Panel operation 2. Press F1 (Marker) repeatedly to select the marker ID from 1 to 5. Marker1 2 3 4 5F 1 3. Press F2 (On) and turn On the selected marker. On OffF 2 4. Make sure Normal is selected in F3. If necessary, press F3 and select it. 5. Repeat the above steps for the number of markers required. Display The upper right corner of the display shows the active marker. Marker ID, Frequency, Amplitude