  • AMPLITUDE 53 1. Press the Amplitude key. Amplitude 2. Press F5 (Corrections). Corrections..F 5 1. Enter correction edit mode 3. Press F1 (Edit). The display shows the correction sets. Edit...F 1 2. Select correction set Press F1 (Select) repeatedly to select the correction set. 5 sets, 30 points each, are selectable. Select1 2 3 4 5F 1 Example: correction set 3 selected Select1 2 3 4 5F 1 3a. Add correction point 1. Make sure that the cursor is pointing to the first empty frequency point. Pt. Freq. (MHz) Gain12345 2. If necessary, move the cursor using the Up/Down keys. 3. Enter the frequency using the Numerical and Unit keys: 9kHz to 3GHz. BK SPdBEnterGHz /SecmSecMHz /kHz / Secμ 4. The cursor automatically moves to the Gain side. Enter the gain using the Numerical keys and dB key Range: −40dB to +40dB. Pt. Freq. (MHz) Gain123452200 BK SPdBEnterGHz /SecmSecMHz /kHz / Secμ 5. Repeat the above procedure for all correction data. The points are automatically sorted by frequency (low → high).