3.0 GHz Spectrum Analyzer GSP830 User Manual

3.0 GHz Spectrum Analyzer GSP830 User Manual
Pages 181
Views 2,164
Downloads : 1 times
PDF Size : 4.4 MiB

Summary of Contents

3.0 GHz Spectrum Analyzer GSP830 User Manual

  • GSP-830 User Manual 160 Command Syntax The commands are partially compatible with IEEE488.2 (1992) and SCPI (Standard Commands for Programmable Instrumentation) (1994) standards. Commands are NON-case sensitive. Example command meas:freq:cen _ 9 _ khz12 3 45 1: Command Header 2: Single space 3: Parameter1 4: Single space 5: Parameter2 Command Header Several command header elements (nodes) can be concatenated to form a complex command. The above example can be separated into: meas: (root node) + freq: + cen: 0/1 0 or 1. Parameter example 1to4 Integer between 1, 2, 3, or 4. 0.01to5 Decimal number between 0.01 and 5. khz Unit (non-case sensitive) Message Terminator Marks the end of a command line. Any of the following is acceptable. CR^END Line feed code (hexadecimal 0D) with END message CR Line feed code <dab>^END Last data byte with END message Message Separator ; (semicolon) Command separator.