3.0 GHz Spectrum Analyzer GSP830 User Manual

3.0 GHz Spectrum Analyzer GSP830 User Manual
Pages 181
Views 1,712
Downloads : 1 times
PDF Size : 4.4 MiB

Summary of Contents

3.0 GHz Spectrum Analyzer GSP830 User Manual

  • REMOTE CONTROL 159 3. Select the GPIB address using the Left/Right keys or Scroll knob. Configure the PC accordingly. 4. Connect the GPIB cable to the rear panel terminal. Check GPIB installation status 1. To check the GPIB installation status, press the System key. System 2. Press F6 (More). More...F 6 3. Press F4 (System Config On). The configuration appears. System ConfigOn OffF 4 4. The system configuration window appears. GPIB sign shows installed (checked) or not installed (unchecked) Installed Not installed Check GPIB self-test result 1. To check the internal GPIB functionality test result, press the System key. System 2. Press F6 (More). More...F 6 3. Press F2 (Self Test). Self Test...F 2 4. The GPIB result appears at F1. If the result is Fail (underlined), contact the service center. GPIBPass FailF 1 GPIB constraints Keep these rules when using the GPIB interface. • Altogether less than 15 devices & 20m cable length, 2m between each device on the bus • Unique address assigned for each device • At least 2/3 of the GPIB devices turned On • No loop or parallel structure allowed Functionality Check Run this query command from the terminal. *idn? This should return the Manufacturer, Model number, Serial number, and Firmware version. GW, GSP-830, P920130, V3.01