3.0 GHz Spectrum Analyzer GSP830 User Manual

3.0 GHz Spectrum Analyzer GSP830 User Manual
Pages 181
Views 1,537
Downloads : 1 times
PDF Size : 4.4 MiB

Summary of Contents

3.0 GHz Spectrum Analyzer GSP830 User Manual

  • GSP-830 User Manual 142 TRACKING GENERATOR The optional Tracking Generator (Opt.01) is a factory installed item which generates a sweep signal with its sweep time and frequency range matching the GSP-830 system. The amplitude is maintained to a constant value over the entire frequency range, which is very useful for testing the frequency response of the DUT. Activate tracking generator 1. Activate TG output 1. Press the Option key. Option 2. Press F1 (TG). TG...F 1 3. Press F1 (TG On). TGOn OffF 1 4. The tracking generator output becomes activated. RF INPUT 50Ω+30dBm MAXMAXDC ±25V 2. Set TG output level 1. Press F2 (TG Level). TG LevelF 2 2. Change the TG output level using the Numerical keys and Enter key. BK SPdBEnterGHz /SecmSecMHz /kHz / Secμ Range 0 to −50dBm Normalize tracking generator 1. Set reference level 1. Press the Option key. Option