3.0 GHz Spectrum Analyzer GSP830 User Manual

3.0 GHz Spectrum Analyzer GSP830 User Manual
Pages 181
Views 1,568
Downloads : 1 times
PDF Size : 4.4 MiB

Summary of Contents

3.0 GHz Spectrum Analyzer GSP830 User Manual

  • TRIGGER 105 Activate external trigger Panel operation 1. Press the Trigger key. Trigger 2. Press F2 (Trigger Condition Ext.). TriggerConditionVideo Ext.F 2 3. The Ext. trigger icon appears at the bottom of the display. 4. Connect the external trigger signal to the rear panel terminal.EXT. TRIG.INPUT Input level range 0V to 5V, positive edge trigger Select Trigger Mode Panel operation 1. Press the Trigger key. Trigger 2. Press F3 (Trigger Mode) repeatedly to select the trigger mode. Trigger ModeNorm. Sgl. Cont.F 3 3. Press F6 (Run Now) to manually start triggering. Run NowF 6 Modes Normal The GSP-830 captures signals every time the triggering condition occurs. Single The GSP-830 captures signals when the first triggering condition occurs, then stops capturing altogether. Continuous The GSP-830 captures signals when the first triggering condition occurs, then switches to Free Run mode. After then, input signals are captured continuously regardless of their conditions.